Triple coincidence positron-lifetime spectrometer



Scheme of triple coincidence positron lifetime spectrometer

  • fast-fast+slow setup
  • detectors: Photonis XP2020/Q photomultipliers equipped cylindrical with BaF2 scintillators
  • electronics: analogue modules NIM (Ortec, Fast-Com Canberra)
  • timing resolution: 220 ps (FWHM 22Na)
  • coincidence count rate: 15/s (4 MBq 22Na)
  • suppression ratio 60Co / 22Na: 5 x 10-3

A detailed description of spectrometer:

J. Čížek, F. Bečvář, I. Procházka: "Three-Detector Setup for Positron-Lifetime Spectroscopy of Solids Containing 60Co Radionuclide", Nuclear Instruments and Methods in Physics A 450, 325-338 (2000).

Our triple coincidence positron lifetime spectrometer was successfully employed for characterization of neutron-irradiated reactor-pressure-vessel steels which contain 60Co radioisotope. Obtained result can be found in the following paper:

J.Kočík, E. Keilová, J. Čížek, I. Procházka: "TEM and PAS study of irradiated reactor pressure vessel steels" Journal of Nuclear Materials 303, 52 (2002)